Diffuse Optical Tomography (DOT) and Imaging (DOI)

Probing Tissue Structure and Function

A quantitative optical imaging technique that probes light propagation in highly scattering media to reconstruct tissue structure and function.
Mammography result obtained using diffuse optical imaging (DOI).
Table of contents

Imaging Deep Tissue with Diffuse Optical Methods

What is DOT & DOI?

Diffuse Optical Tomography (DOT) and Diffuse Optical Imaging (DOI) are non-invasive optical techniques that use near-infrared light to probe the optical properties of biological tissue. They are particularly well suited for soft, highly scattering tissues such as brain, breast, and muscle.

By measuring how light is absorbed and scattered while propagating through tissue, DOT and DOI enable the reconstruction of spatial maps of physiological parameters, including hemoglobin concentration, blood oxygenation, and tissue scattering. These parameters provide insight into tissue composition, perfusion, and functional activity at depths inaccessible to conventional optical microscopy.

How does DOT & DOI work?

In Diffuse Optical Tomography (DOT) and Diffuse Optical Imaging (DOI), near-infrared light is delivered as short laser pulses or as modulated continuous waves and introduced into biological tissue at multiple positions. As photons propagate through the tissue, they undergo multiple scattering events and partial absorption by intrinsic chromophores such as oxyhemoglobin, deoxyhemoglobin, lipids, and water.

The emerging light, which has become temporally broadened and attenuated through these interactions, is detected at several surface locations using time-resolved or frequency-domain detection schemes. By recording the temporal or phase-resolved distribution of photon arrival times and by applying physical models of photon transport, the tissue’s absorption and scattering coefficients can be estimated. These optical parameters provide the basis for reconstructing three-dimensional maps that represent tissue composition and functional activity.

In fluorescence-assisted implementations, an additional contrast agent is introduced. Fluorescence photons emitted by the agent are detected together with the transmitted light, which allows quantitative localization and characterization of fluorophores or targeted probes within deep tissue.

DOT & DOI Data & Analysis

Time-resolved DOT and DOI measurements generate temporal point spread functions (TPSFs), which represent the distribution of photon arrival times and describe how photons propagate through scattering tissue. These datasets are analyzed with model-based reconstruction algorithms that solve the inverse problem of estimating tissue optical properties from the measured photon distributions.

By fitting the recorded temporal profiles with diffusion or radiative transfer models, quantitative three-dimensional maps of optical absorption and scattering coefficients can be obtained. From these parameters, hemoglobin concentration and oxygen saturation are derived. In fluorescence-based DOI, additional analysis provides the spatial localization and concentration of fluorescent contrast agents within the tissue volume.

Why use DOT & DOI?

Diffuse Optical Tomography (DOT) and Diffuse Optical Imaging (DOI) provide unique access to physiological information in deep tissue using near-infrared light, without the need for ionizing radiation or invasive procedures. They enable functional imaging of blood oxygenation, perfusion, and microvascular hemodynamics, making them well suited for longitudinal studies and clinical research.

Because DOT and DOI are sensitive to both optical absorption and scattering, they can detect changes in tissue composition and function that are not accessible with purely anatomical imaging techniques. This capability makes them valuable tools for applications such as functional brain imaging, breast cancer research, stroke monitoring, and therapy assessment.

Instrumentation requirements for DOT & DOI

Reliable diffuse optical tomography and imaging require time-resolved excitation and detection with high temporal precision and multi-channel acquisition capability. Key instrumentation requirements include:

  • Picosecond pulsed laser sources operating in the near-infrared range with stable repetition rates for controlled tissue excitation.
  • Single-photon sensitive detectors, such as photomultiplier tubes or single-photon avalanche diodes, optimized for near-infrared wavelengths.
  • Multi-channel TCSPC electronics that record photon arrival times with picosecond resolution across multiple source and detector pairs.

Together, these components enable precise characterization of photon propagation in scattering tissue and provide the foundation for quantitative reconstruction of physiological parameters.

Relevant for Your Research​

Matching Applications

Fluorescence lifetime imaging microscopy of plant tissue showing structural organization through intrinsic autofluorescence from plant cell components.
Life Science
Fluorescence lifetime imaging of MDCK cell membranes labeled with the tension probe Flipper-TR visualizes membrane tension changes during osmotic stress in mechanobiology experiments.
Life Science

PicoQuant Instrumentation for DOT & DOI

Explore PicoQuant’s compatible components designed for DOT & DOI.

PicoQuant pulsed laser and LED sources

Laser Sources

Picosecond pulsed diode lasers provide controlled near-infrared excitation for time-resolved DOT and DOI measurements. Stable repetition rates and short pulse durations enable precise characterization of photon propagation in scattering tissue.

PicoQuant's hybrid and SPAD single-photon detectors

Single-Photon Counting Detectors

Single-photon sensitive detectors record weak transmitted or fluorescent signals with high timing precision. Low timing jitter and high sensitivity are essential for accurate extraction of absorption and scattering properties.

Overview image of PicoQuant Time Tagging and TCSPC units including HydraHarp 500, MultiHarp 150/160, PicoHarp 330, and TimeHarp 260.

Time Tagging & TCSPC Electronics

Time-Correlated Single Photon Counting (TCSPC) and time tagging electronics measure photon arrival times with picosecond resolution across multiple channels. This enables precise reconstruction of temporal point spread functions for quantitative tissue imaging.

Info Request

Info Request

Contact us

Please fill out the form below to request more information about our products and services. You may also use it to ask for pricing, availability, technical specifications, or any other details relevant to your inquiry. Our team will be happy to review your request and get in touch with you. If additional information is needed to process your inquiry, we will let you know.

* Required