Lifetime upgrade kit for PTI QuantaMaster |
QM-Upgrade Kit |
Specifications
|
Detectors |
|
|
| |
Type* |
|
PMA Photomultiplier assembly |
| |
Spectral range |
|
185 - 650 nm |
 |
185 - 820 nm |
| |
Dark counts (20°C, typ. value) |
|
< 50 cps |
 |
< 900 cps |
| |
Transit time spread (FWHM, typ. value) |
|
200 ps |
 |
| |
TCSPC Data Acquisition |
|
|
| |
Type |
|
TimeHarp 200 |
 |
PicoHarp 300 |
 |
NanoHarp 250 |
| |
Time resolution (bin width) |
|
< 40 ps |
|
4 ps |
|
4 ns or 32 ns (selectable) |
| |
Dead time |
|
< 350 ns |
|
< 95 ns |
|
< 8 ns |
| |
Number of time channels |
|
up to 4096 |
|
up to 65536 |
|
up to 262144 |
| |
Time range |
|
150 ns to 4.5 µs |
|
260 ns to 33 µs |
|
up to 1.04 ms resp. 2.15 s (depending on selected time resolution) |
 |
| |
Software features (FluoFit) |
|
|
| |
Fitting algorithm |
 |
nonlinear least squares fitting (Marquardt-Levenberg, Monte Carlo, manual parameter variation), iterative reconvolution, global fitting for all models |
| |
Fitting models |
 |
exponential decay up to 4th order, lifetime distributions (Gaussian, Lorentzian, Stretched Exponential), anisotropy (up to 4th order exponential decay model) |
| |
Error analysis |
|
χ2, distribution and autocorrelation of weighted residuals, Asymptotic Standard Errors (ASE), Support Plane Analysis and Bootstrap |
 |
| |
Operational & Electrical |
|
|
| |
Operating environment |
|
PC Pentium class, 2 GHz, 512 MB memory with Windows 2000/XP |
| |
Power requirements |
|
110/230 V, 50/60 Hz |
 |
| * lasers, different detectors and cooling available upon request |
 |
All Information given here is reliable to our best knowledge. However, no responsibility is assumed for possible inaccuracies or omissions. Specifications and external appearances are subject to change without notice.
© PicoQuant GmbH, 1998 - 2012
|