Compact Lifetime and FCS Upgrade Kit for LSMs LSM Upgrade Kit 

Specifications

Excitation Source*  
Light source   Laser Diode Heads (LDH Series)
Wavelengths   405 - 510 nm, 530 nm, 635 - 900 nm (375 nm upon request)
Repetition rate   up to 40 MHz (optional 80 MHz)
Detectors  
Type PMT (PMA Series) SPAD (τ-SPAD) SPAD (PDM Series)
Spectral range* 185 - 650 nm, 185 - 820 nm 400 - 1000 nm 400 - 1000 nm
Dark counts (at 20°C, typ. value) < 50 cps < 900 cps < 100 cps < 250 cps
Instrument Response Function
(IRF at 650 nm)
typ. 200 ps typ. 400 ps typ. 50 ps
TCSPC Data Acquisition  
Type   PicoHarp 300
Time resolution (bin width)   4 ps
Dead time   < 95 ns
Time range   260 ns to 2 µs
Time tag resolution (TTTR mode)   equals excitation period
Sustained data throughput in TTTR mode   up to 5×106 counts/sec
Maximum collection time   virtually unlimited, restricted only by computer memory/hard disk capacity
Software Features (SymPhoTime)  
General concept   use of versatile TTTR file format for data acquisition, data archiving in workspace, time gating for all methods, separation of up to four detector signals
Point measurements   data conversion to: MCS trace, FCS calculation and fitting, TCSPC histogram, on/off-state histogram, burst size analysis, (PIE-)FRET histogram, photon counting histogram, lifetime histogram, FCCS, FLCS
Fluorescence Lifetime Imaging (FLIM)   data conversion to: fluorescence intensity images (max. image size 512 × 512 pixel), fluorescence lifetime images, time gated analysis, TCSPC histogram for region of interest
Operational & Electrical  
Operating environment   PC Pentium/Athlon class, 3 GHz, dual core, 4 GB memory with Windows 7
Power requirements   110/230 V, 50/60 Hz
* lasers, different detectors and cooling available upon request

All Information given here is reliable to our best knowledge. However, no responsibility is assumed for possible inaccuracies or omissions. Specifications and external appearances are subject to change without notice.
© PicoQuant GmbH, 1998 - 2012
back to top

Home PicoQuant GmbH
Rudower Chaussee 29 (IGZ)
12489 Berlin
Germany
Phone:  ++4930 6392 6929
Fax: ++4930 6392 6561
Email: infopicoquant.com