Ultraviolet to Infrared Picosecond Diode Laser Heads |
LDH 375 to 1550 |
Laser Head Specifications (typical values)
| Beam Parameters |
|
| Optics focus length |
f' = 4.5 mm |
| Numerical aperture |
0.55 |
| Typical divergence |
Theta parallel 0.32 mrad |
| (with optics) |
Theta perpendicular 0.11 mrad |
| Polarisation |
typ. linear, perpendicular to the longer axis of the elliptical beam |
| Polarisation degree |
> 90% |
| Sidemode suppression ratio (SMSR) |
<0.01 |
| |
|
| Cooling |
|
| Peltier cooling stability |
better than 1 K for ambient temperature between 15°C and 30°C |
| |
|
| Dimensions |
|
| Cooled |
110 × 75 × 60 mm (l × w × h) - See layout |
| Cooled with fibre coupling |
145 × 75 × 60 mm (l × w × h) - See layout |
| Uncooled |
25 × 76 mm (diameter × length with collimator) - See layout |
| Uncooled with fibre coupling |
25 × 106 mm (diameter × length with collimator) - See layout |
| |
|
| Spectral width |
approx. 2-8 nm for wavelengths < 900 nm |
| |
approx. 10-20 nm for wavelengths > 900 nm |
| |
<1 nm in cw operation |
| |
|
| Power stability (cooled) |
|
| 12 hour, Delta T (ambient) <3 K |
1 % RMS |
| |
3 % Peak to Peak |
 |
Beam Shape and Fibre Coupling
All laser heads emit a collimated elliptical beam with typical dimensions of 1.5 × 3.5 mm. They can be optionally fitted with single mode, polarisation maintaining single mode or multi mode optical fibres through appropriate fibre couplers. Different fibre connectors at the fibre output are available (FC/PC, FC/APC and SMA). For details, please see the corresponding webpage.

Available wavelengths (pulsed mode, LDH-P Series)
These laser heads can only be driven either by the PDL 800-B, PDL 800-D, PDL 808 "Sepia" or PDL 828 "Sepia II" driver units.
| |
|
Power Adjust = Low (Narrow Pulse) |
Power Adjust = High (Wide Pulse) |
Wavelength
nm ( ± 10) |
Type |
Pulse FWHM
ps |
Average Power
mW (@ 40 MHz) |
Average Power
mW (@ 80 MHz) |
Pulse FWHM
ps |
Average Power
mW (@ 40 MHz) |
Average Power
mW (@ 80 MHz) |
| 375 |
LDH-P-C-375 |
< 70 |
0.3 |
-- |
< 300 |
1.0 |
-- |
| 375 |
LDH-P-C-375B |
< 70 |
0.3 |
-- |
< 300 |
3.0 |
-- |
| 405 |
LDH-P-C-405 |
< 70 |
0.4 |
0.8 |
< 300 |
2.0 |
4.0 |
| 405 |
LDH-P-C-405B |
< 70 |
1.0 |
-- |
< 300 |
3.0 |
-- |
| 405 |
LDH-P-C-405M |
< 70 |
2.5 |
-- |
< 600 |
20.0 |
-- |
| 440 |
LDH-P-C-440 |
< 90 |
0.3 |
-- |
< 300 |
1.0 |
-- |
| 440 |
LDH-P-C-440B |
< 70 |
0.3 |
-- |
< 500 |
3.0 |
-- |
| 440 |
LDH-P-C-440M |
< 70 |
3.0 |
-- |
< 500 |
20.0 |
-- |
| 450 |
LDH-P-C-450B |
< 70 |
0.7 |
-- |
< 300 |
5.0 |
-- |
| 470 |
LDH-P-C-470 |
< 90 |
0.3 |
-- |
< 300 |
1.0 |
-- |
| 470 |
LDH-P-C-470B |
< 90 |
0.3 |
-- |
< 500 |
3.0 |
-- |
| 485 |
LDH-P-C-485 |
< 90 |
0.3 |
-- |
< 500 |
2.0 |
-- |
| 485 |
LDH-P-C-485B |
< 100 |
0.6 |
-- |
< 550 |
5.0 |
-- |
| 531 |
LDH-P-FA-530 |
Please see website for LDH-P-FA-530 |
| 635 |
LDH-P-635 |
< 90 |
0.3 |
0.6 |
< 400 |
2.5 |
4.0 |
| 635 |
LDH-P-C-635B |
< 90 |
1.5 |
3.0 |
< 300 |
10.0 |
20.0 |
| 640 |
LDH-P-C-640B |
< 90 |
0.7 |
1.5 |
< 400 |
4.5 |
9.0 |
| 655 |
LDH-P-650 |
< 70 |
0.4 |
0.9 |
< 400 |
3.0 |
6.0 |
| 660 |
LDH-P-660 |
< 90 |
0.5 |
1.0 |
< 500 |
5.0 |
10.0 |
| 670 |
LDH-P-670 |
< 70 |
0.3 |
0.7 |
< 300 |
1.0 |
2.0 |
| 665 |
LDH-P-670B |
< 90 |
0.3 |
0.7 |
< 400 |
2.0 |
4.0 |
| 690 |
LDH-P-690 |
< 70 |
0.5 |
1.0 |
< 400 |
4.0 |
8.0 |
| 705 |
LDH-P-705 |
< 70 |
1.2 |
2.5 |
< 400 |
5.0 |
10.0 |
| 735 |
LDH-P-730 |
< 130 |
0.3 |
0.6 |
< 400 |
1.7 |
3.5 |
| 757 |
LDH-P-760 |
< 90 |
0.4 |
0.9 |
< 300 |
3.0 |
6.0 |
| 785 |
LDH-P-780 |
< 70 |
0.5 |
1.0 |
< 500 |
5.0 |
10.0 |
| 806 |
LDH-P-810 |
< 90 |
0.4 |
0.9 |
< 500 |
5.0 |
10.0 |
| 830 |
LDH-P-830 |
< 70 |
0.3 |
0.6 |
< 300 |
4.0 |
8.0 |
| 835 |
LDH-P-840 |
< 90 |
1.0 |
2.0 |
< 750 |
6.0 |
12.0 |
| 850 |
LDH-P-850 |
on request
|
| 905 |
LDH-P-905 |
< 90 |
0.5 |
1.0 |
< 300 |
4.0 |
8.0 |
| 952 |
LDH-P-950 |
< 90 |
0.2 |
0.4 |
< 400 |
2.0 |
4.0 |
| 965 |
LDH-P-965 |
< 90 |
1.2 |
2.5 |
< 400 |
10.0 |
20.0 |
| 975 |
LDH-P-980** |
< 90 |
0.6 |
1.2 |
< 750 |
10.0 |
20.0 |
| 1025 |
LDH-P-1025 |
on request
|
| 1060 |
LDH-P-1060 |
< 90 |
0.5 |
1.0 |
< 550 |
10.0 |
20.0 |
| 1080 |
LDH-P-C-1080* |
|
< 50 |
2.5 |
5.0 |
| 1120 |
LDH-P-C-1120* |
|
< 90 |
2.5 |
5.0 |
| 1310 (± 20 nm) |
LDH-P-1310 |
< 50 |
0.08 |
0.15 |
< 400 |
1.0 |
2.0 |
| 1550 (± 30 nm) |
LDH-P-1550 |
< 50 |
0.01 |
0.02 |
< 400 |
0.5 |
1.0 |
| 1550 |
LDH-P-F-1550 |
< 110 |
0.3 |
0.7 |
< 500 |
4.0 |
8.0 |
*pulse width independent from selected power level
**not suited for fibre coupling
Option "B": High Power version Option "C": Laser head with thermoelectric cooler, mandatory for some laser heads, optional for all other laser heads Option "M": Transversal multi mode structure, reduced coupling efficiency into single mode fibres Option "F": Laser head emitting divergent beam from FC/APC fibre coutput connector
All measurements may be subject to a 10% calibration error.
The two power adjustment levels specified here refer to the same laser head. These levels can be adjusted using the corresponding driver (PDL 800-B, PDL 800-D, PDL 808 or PDL 828). The 'Low' adjustment is the best choice for shortest pulses and is usually reached close to the lasing threshold. The 'High' adjustment is used to achieve highest pulse power at moderate pulse length and corresponds to the maximum intensity setting of the driver.

Available wavelengths (dual mode: cw and pulsed, LDH-D Series)
These laser heads can only be driven by the PDL 800-D or the PDL 828 "Sepia II" driver.
| |
|
|
Power Adjust = Low (Narrow Pulse) |
Power Adjust = High (Wide Pulse) |
Wavelength
nm ( ± 10) |
Type |
CW Power mW |
Pulse FWHM
ps |
Average Power
mW (@ 40 MHz) |
Average Power
mW (@ 80 MHz) |
Pulse FWHM
ps |
Average Power
mW (@ 40 MHz) |
Average Power
mW (@ 80 MHz) |
| 375 |
LDH-D-C-375 |
10 |
< 70 |
0.3 |
-- |
< 300 |
3.0 |
-- |
| 405 |
LDH-D-C-405 |
20 |
< 70 |
1.0 |
-- |
< 300 |
3.0 |
-- |
| 405 |
LDH-D-C-405M |
400 |
< 70 |
2.5 |
-- |
< 600 |
20.0 |
-- |
| 440 |
LDH-D-C-440 |
20 |
< 70 |
0.3 |
-- |
< 500 |
3.0 |
-- |
| 440 |
LDH-D-C-440M |
400 |
< 70 |
3.0 |
-- |
< 500 |
20.0 |
-- |
| 450 |
LDH-D-C-450 |
10 |
< 70 |
0.7 |
-- |
< 300 |
5.0 |
-- |
| 470 |
LDH-D-C-470 |
10 |
< 90 |
0.3 |
-- |
< 500 |
3.0 |
-- |
| 485 |
LDH-D-C-485* |
10 |
< 90 |
0.3 |
-- |
< 500 |
2.0 |
-- |
| 635 |
LDH-D-C-635 |
50 |
< 90 |
1.5 |
3.0 |
< 300 |
10.0 |
20.0 |
| 635 |
LDH-D-C-635M |
100 |
< 90 |
1.0 |
2.0 |
< 400 |
4.5 |
9.0 |
| 640 |
LDH-D-C-640 |
20 |
< 90 |
0.7 |
1.5 |
< 400 |
4.5 |
9.0 |
| 660 |
LDH-D-C-660 |
30 |
< 70 |
0.7 |
1.5 |
< 500 |
4.5 |
9.0 |
| 670 |
LDH-D-C-670 |
15 |
< 70 |
0.3 |
0.6 |
< 600 |
4.0 |
8.0 |
| 690 |
LDH-D-C-690 |
20 |
< 70 |
0.5 |
1.0 |
< 500 |
4.0 |
8.0 |
| 730 |
LDH-D-C-730 |
8 |
< 120 |
0.3 |
0.7 |
< 800 |
1.7 |
3.5 |
| 1080 |
LDH-D-C-1080** |
30 |
|
< 50 |
2.5 |
5.0 |
| 1120 |
LDH-D-C-1120** |
20 |
|
< 90 |
2.5 |
5.0 |
| other wavelengths are available upon request |
*different coupling efficiency into optical fibres for pulsed and cw operation due to astigmatism
**pulse width independent from selected power level
Option "C": Laser head with thermoelectric cooler Option "M": Transversal multi mode structure, reduced coupling efficiency into single mode fibres
All measurements may be subject to a 10% calibration error.
The two power adjustment levels specified here refer to the same laser head. These levels can be adjusted using the corresponding driver (PDL 800-D or PDL 828). The 'Low' adjustment is the best choice for shortest pulses and is usually reached close to the lasing threshold. The 'High' adjustment is used to achieve highest pulse power at moderate pulse length and corresponds to the maximum intensity setting of the driver.

Available wavelengths - narrow spectral bandwidth
These laser heads can only be driven either by the PDL 800-B, PDL 800-D, PDL 808 "Sepia" or PDL 828 "Sepia II" driver units.
| Wavelengths |
Power Adjust = Low (Narrow Pulse) |
|
Power Adjust = High (Wide Pulse) |
Central
nm |
Variation
nm |
Spectral FWHM nm |
Pulse FWHM
ps |
Average Power
mW (@ 40 MHz) |
Average Power
mW (@ 80 MHz) |
Pulse FWHM
ps |
Average Power
mW (@ 40 MHz) |
Average Power
mW (@ 80 MHz) |
| 772 |
±3 |
<1 |
<120 |
1.2 |
2.5 |
<350 |
3.5 |
7.0 |
| 784 |
±3 |
<0.3 |
<100 |
0.9 |
1.6 |
<500 |
4.0 |
8.0 |
| 852 |
±3 |
<0.3 |
<90 |
2.0 |
3.5 |
<350 |
5.0 |
9.0 |
| 976* |
±3 |
<0.5 |
--- |
--- |
--- |
<90 |
5.0 |
10.0 |
| 1062 |
±3 |
0.4 - 0.8 |
<120 |
0.7 |
1.4 |
<400 |
5.0 |
10.0 |
| 1275 |
±7 |
0.2 |
<100 |
0.01 |
0.03 |
<400 |
0.5 |
1.0 |
| 1300 |
±7 |
0.2 |
<100 |
0.01 |
0.03 |
<400 |
0.5 |
1.0 |
| 1310 |
±20 |
0.2 |
<100 |
0.01 |
0.03 |
<400 |
0.5 |
1.0 |
| 1325 |
±7 |
0.2 |
<100 |
0.01 |
0.03 |
<400 |
0.5 |
1.0 |
| 1349 |
±7 |
0.2 |
<100 |
0.01 |
0.03 |
<400 |
0.5 |
1.0 |
| 1417 |
±3 |
0.2 |
<100 |
0.01 |
0.03 |
<400 |
0.5 |
1.0 |
| 1490 |
±3 |
0.2 |
<100 |
0.01 |
0.03 |
<400 |
0.5 |
1.0 |
| 1510 |
±3 |
0.2 |
<100 |
0.01 |
0.03 |
<400 |
0.5 |
1.0 |
| 1530 |
±3 |
0.2 |
<100 |
0.01 |
0.03 |
<400 |
0.5 |
1.0 |
| 1550 |
±3, ±20 |
0.2 |
<100 |
0.01 |
0.03 |
<400 |
0.5 |
1.0 |
| 1570 |
±3 |
0.2 |
<100 |
0.01 |
0.03 |
<400 |
0.5 |
1.0 |
| 1590 |
±3 |
0.2 |
<100 |
0.01 |
0.03 |
<400 |
0.5 |
1.0 |
| 1610 |
±3 |
0.2 |
<100 |
0.01 |
0.03 |
<400 |
0.5 |
1.0 |
| Available upon request: 760, 763, 780, 795, 937, 1083 nm (± 3 nm) |
*pulse width independent from selected power level
All measurements may be subject to a 10% calibration error. Please call for details and availability.
Available wavelengths - short pulses (on selection)
These laser heads can only be driven either by the PDL 800-B, PDL 800-D, PDL 808 "Sepia" or PDL 828 "Sepia II" driver units.
Wavelength
nm (+/-10) |
typ. pulse FWHM
ps |
Average Power
mW (@ 40 MHz) |
Average Power
mW (@ 80 MHz) |
| 655 |
54 |
0.4 |
0.9 |
| 670 |
39 |
0.3 |
0.7 |
| 690 |
54 |
0.5 |
1.0 |
| 785 |
54 |
0.5 |
1.0 |
| 830 |
44 |
0.3 |
0.7 |
All measurements may be subject to a 10% calibration error. Please call for details and availability.
Pulse-to-pulse jitter
| The image on the right shows a typical measurement result of the pulse-to-pulse jitter. In this case the measurement result is 2.6 ps. |
 |
A range of sub-nanosecond pulsed LEDs is also available e.g. for time-resolved fluorescence experiments not requiring coherence.
All Information given here is reliable to our best knowledge. However, no responsibility is assumed for possible inaccuracies or ommisions. Specifications and external appearences are subject to change without notice. © PicoQuant GmbH, 1998 - 2009
last updated 17.02.2010
|