Multichannel Picosecond Event Timer & TCSPC Module   HydraHarp 400 

Specifications

 
Input Channels and Sync Constant Fraction Discriminator on all channels, software adjustable
Input voltage range 0 to -800 mV, optimum: -200 mV to -400 mV
Trigger point Falling edge
Trigger pulse width 0.5 to 30 ns
Trigger pulse rise/fall time 2 ns max.
External Reference Clock
Input 10 MHz, min. 200 mV, max. 1 V pp, 50 Ohms, AC coupled
Output 10 MHz, 300 mV pp, 50 Ohms, AC coupled
Time to Digital Converters
Minimum time bin width 1 ps
Electrical time resolution < 12 ps rms
Full scale range - Histogram mode 65 ns to 65 μs (depending on chosen resolution: 1, 2, 4, ..., 1024 ps)
Full scale range - Time-Tagged mode Infinite
Maximum count rate per input channel 12.5×106 counts/sec
Maximum sync rate 150 MHz
Adjustable delay range for each input channel ± 100 ns, resolution 1 ps
Dead time < 80 ns
Differential non-linearity < 2% peak, < 0.2% rms (over full measurement range)
Histogrammer
Count depth per time bin 4.294.967.296 (32 bit)
Maximum number of time bins 65536
Acquisition time 1 ms to 100 hours
TTTR Engine
T2 mode resolution 1 ps
T3 mode resolution 1, 2, 4, ..., 1024 ps
FiFo buffer depth (records) 2097152
Sustained throughput typ. 9×106 events/sec
Operation
PC interface USB 2.0 high speed
PC requirements min. 1 GHz CPU clock, 512 MB memory
Operating system Windows™ 2000/XP/Vista
Power consumption max. 50 W at 100 to 240 VAC

All Information given here is reliable to our best knowledge. However, no responsibility is assumed for possible inaccuracies or ommisions. Specifications and external appearences are subject to change without notice.
© PicoQuant GmbH, 1998 - 2008

last updated 08.05.2008

Home PicoQuant GmbH
Rudower Chaussee 29 (IGZ)
12489 Berlin
Germany
Phone:  ++4930 6392 6560
Fax: ++4930 6392 6561
Email: infopicoquant.com
WWW: http://www.picoquant.com