Compact Time-resolved Spectrometer |
FluoTime 100 |
Specifications
| System |
| Optical configuration |
 |
L-Geometry |
| Mode of operation |
 |
Time-Correlated Single Photon Counting (TCSPC) |
 |
| Excitation Sources |
| Light source |
 |
Pulses LEDs (PLS) |
 |
Laser Diode Heads (LDH) |
| Wavelengths |
|
260 nm to 600 nm |
|
375 - 485 nm, 635 - 1550 nm |
| Pulse width |
|
400 ps - 1 ns |
|
60 ps - 500 ps |
| Repetition rate |
|
Up to 40 MHz |
|
Up to 40 MHz (80 MHz optional) |
 |
| Detectors |
| Type |
 |
PMA Photomultipier Assembly |
| Spectral Range |
|
185 - 650 nm |
185 - 820 nm |
300 - 850 nm |
| Dark Counts (at 25°C) |
|
< 100 cps |
< 1000 cps |
< 2000 cps |
|
|
other detectors and cooling available upon request |
 |
| Data Acquisition Board |
| Type |
 |
NanoHarp 250 |
 |
PicoHarp 300 |
| Number of time channels/curve |
|
up to 262144 |
|
65536 |
| Count depth |
|
18 bit |
|
16 bit |
| Time resolution (bin width) |
|
4 ns or 32 ns |
|
4 ps |
| Lifetime resolution |
|
< 10 ns |
|
< 40 ps |
| Differential non-linearity |
|
< 0.1% rms, < 1% pp. |
|
< 1% rms, < 5% pp |
| Dead time |
|
8 ns |
|
< 95 ns |
| Discriminator level, zero cross adjust |
|
Software adjustable |
| Collection time |
|
1 ms to 1 h |
 |
| Operation Environment |
| Operating system |
 |
Windows® 2000/XP |
| Power consumption |
|
220/240 or 110/120 VAC, 50/60 Hz |
| Dimensions |
|
300 mm × 200 mm × 100 mm (w × d × h) |
All Information given here is reliable to our best knowledge. However, no responsibility is assumed for possible inaccuracies or ommisions. Specifications and external appearences are subject to change without notice.
© PicoQuant GmbH, 1998 - 2006
|