Compact time resolved spectrometer FluoTime 100 

Measurement Examples


Coumarin 152 in EtOH

  • Time-resolved fluorescence lifetime measurement of Coumarin 152 (3·10-7 molar) in Ethanol using the FluoTime 100, TimeHarp 200 and PMT module PMA 182.
  • Excitation with LDH-P-C-405 picosecond diode laser, 50 ps FWHM at 20 MHz repetition rate.
  • The plot shows the measured Instrument Response Function (red), the sample decay (blue) and the fitted decay. The fluorescence lifetime is 1.59 ns. An excellent chi-square of 0.912 and the residual trace demonstrate the quality of the fit. 



Oxazin 1 in H2O

  • Time-resolved fluorescence lifetime measurement of Oxazin 1 (1·10-6 molar) in Water using the FluoTime 100, TimeHarp 200 and PMT module PMA 182.
  • Excitation at 635 nm with LDH-P-635 picosecond diode laser, 70 ps FWHM at 20 MHz repetition rate.
  • The plot shows the measured Instrument Response Function (red), the sample decay (blue) and the fitted decay. The fluorescence lifetime is 799 ps. An excellent chi-square of 1.051 and the residual trace demonstrate the quality of the fit.



Oxazin 1 and Oxazin 4 mixture in Ethanol

  • Fluorescence decay measurement of Oxazine 1 with a trace amount of Oxazine 4 (10-6 molar solution in Ethanol). Data acquired using a FluoTime 100 spectrometer, TimeHarp 200 and a LDH-P-635.
  • Fluorescence with λ>670 nm was collected. The plot shows the measured Instrument Response Function (red), sample decay (blue) and the fitted double exponential curve (black). The two recovered lifetimes are 0.97 ± 0.02 ns and 3.71 ± 0.09 ns, respectively. They are in excellent agreement with the single exponential lifetimes of Oxazine 1 and Oxazine 4, respectively, determined in independent experiments. The weighted residuals' trace and chi-square = 1.158 demonstrates the quality of the fit.

All Information given here is reliable to our best knowledge. However, no responsibility is assumed for possible inaccuracies or ommisions. Specifications and external appearences are subject to change without notice.