Dual SPAD Power Supply DSN 101 

Specifications

Detector Input
Pulse width >20 ns
Pulse height >2.3 V
Electrical Parameters
High shut-down count rate factory set:
- PDM SPADs: 8×106counts/sec
- SPCM-AQR(H) SPADs: 4×106counts/sec
Oscilloscope output max. 2 V
Acoustic output >120 Ohms
Connectors
SPAD power socket LEMO EGG 2B 305 CLL
Appropriate male connector LEMO FGG 2B 305 CLL
Input / output SPAD female SMA
Oscilloscope output female BNC
Acoustic output 3.5 mm earphone socket
Power Requirements Stand alone version OEM version
Voltage 110 to 240 V
50/60 Hz
+5 V
stabilized at ± 5%
Power / Current consumption max. 35 Watts max. 3 A
Dimensions Stand alone version
(including rack)
OEM version
(3U / 21 HP)
Width 237 mm 106 mm
Depth 310 mm 173 mm
Height 133 mm 129 mm

All Information given here is reliable to our best knowledge. However, no responsibility is assumed for possible inaccuracies or omissions. Specifications and external appearances are subject to change without notice.